1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
395
User Manual 1173.9511.02 ─ 12
{<2_SegReliability>
0 | 3 | 4 | 8
The segment reliability indicates whether one of the following
exceptions occurred in this segment:
0:
No error
3:
Signal overflow
4:
Signal low
8:
Synchronization error
If a combination of exceptions occurs, the most severe error is
indicated.
<3_RRI>, <4_Pilot>,
<5_ACKDSC>,
<6_AuxPilot>,
<7_DRC>, <8_Data>}
RMS channel power values for the physical channels. Depend-
ing on the subtype 0/1 or 2 some channels exist or not. If not
available, no power measurement is possible therefore
NAV
is
returned.
Range:
-60 dB to +10 dB
Default unit: dB
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V2.0.20
Options:
R&S CMW-KM012
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
FETCh:EVDO:MEAS<i>:MEValuation:LIST:CP:STATe?
Return the states of the channels for power measurement (CP).
The values listed below in curly brackets {} are returned for each active segment:
{...}
seg 1
, {...}
seg 2
, ..., {...}
seg n
. The number of active segments n is determined by
CONFigure:EVDO:MEAS<i>:MEValuation:LIST:COUNt
.
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Return values:
<1_Reliabiltiy>
Conventions and General Information
. In list mode, a zero relia-
bility indicator indicates that the results in all measured seg-
ments are valid. A non-zero value indicates that an error occur-
red in at least one of the measured segments.
Command Reference