CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
785
User Manual 1173.9511.02 ─ 12
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com
●
104 (Not Functional)
:
The application could not be started with the configured parameter set.
5.5.2 General Measurement Settings
The commands valid for all CDMA2000 TX measurements are divided into the groups
listed below.
5.5.2.1
Signal Routing
The following commands configure the scenario, select the connector and define an
external attenuation value.
.............................................................................................. 785
............................................................................... 786
ROUTe:CDMA:MEAS<i>:SCENario:SALone
.....................................................................786
ROUTe:CDMA:MEAS<i>:SCENario:CSPath
.....................................................................786
ROUTe:CDMA:MEAS<i>:SCENario:CATalog:CSPath?
......................................................787
CONFigure:CDMA:MEAS<i>:RFSettings:EATTenuation
.................................................... 787
ROUTe:CDMA:MEAS<i>?
Returns the configured routing settings.
For possible connector and converter values, see
chapter 3.5.1.4, "Values for RF Path
Return values:
<Scenario>
SALone | CSPath
SALone
: Standalone (Non Signaling)
CSPath
: Combined Signal Path
<Controller>
Controlling application for scenario CSPath
<RXConnector>
RF connector for the input path
<RXConverter>
RX module for the input path
Usage:
Query only
Firmware/Software:
V2.0.10
Manual operation:
See
Command Reference