1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
130
User Manual 1173.9511.02 ─ 12
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
To restore these settings, restart your instrument and press the "Factory
Default" softkey during startup.
–
If a software package (update) has not been properly installed this is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
A software update correcting the error may be available. Updates are e.g. pro-
vided in the "CMW Customer Web" on GLORIS (registration required):
.
If you get firmware errors even with the properly installed latest software version,
please send a problem report including log files to Rohde & Schwarz.
●
102 (Unidentified Error)
:
Indicates an error not covered by other reliability values. For troubleshooting
please follow the steps described for "101 (Firmware Error)".
●
103 (Parameter Error)
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com
●
104 (Not Functional)
:
The application could not be started with the configured parameter set.
In the 1xEV-DO signaling application, reliability indicators are returned with RX mea-
surement results - for the following more specific reasons:
Reliability Indicator
Reason
6
A Test Application report was not received from the mobile within 3 update
time periods.
19
Not in connected state.
103
●
The 1xEVDO signaling Test Application is not enabled
●
Forward Link measurement data were requested, but the Test Applica-
tion direction is not set to "Forward" or "Forward & Reverse"
●
Reverse Link measurement data were requested, but the Test Applica-
tion direction is not set to "Reverse" or "Forward & Reverse"
●
Measurement data were requested, but the corresponding measure-
ment is not assigned