1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
284
User Manual 1173.9511.02 ─ 12
Measure on Exception
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected. A faulty result occurs e.g. when an overload is detected. In remote
control, the cause of the error is indicated by the "reliability indicator".
●
Off:
Faulty results are rejected. The measurement is continued; the statistical
counters are not re-set. Use this mode to ensure that a single faulty result does not
affect the entire measurement.
●
On:
Results are never rejected. Use this mode e.g. for development purposes, if
you want to analyze the reason for occasional wrong transmissions.
Remote command:
CONFigure:EVDO:MEAS<i>:OLTR:MOEXception
3.3.13 OLTR: Limits
The OLTR-specific "Limits" can be set in the configuration dialog of the OLTR mea-
surement.
Fig. 3-31: OLTR limits
Initial Lower Limit
Sets the lower limit for the limit violation calculation. Softkey is also available.
Remote command:
CONFigure:EVDO:MEAS<i>:OLTR:LIMit:ILOWer
3.3.14 Generator Shortcut
The "Generator Shortcut " parameters configure the GPRF connection for both, the
multi evaluation and OLTR measurements.
For the parameter description refer to
chapter 3.3.10, "Generator Shortcut"
3.3.15 OLTR Measurement Results
The results of the 1xEV-DO OLTR measurement are displayed in several different
views; see detailed description in section
chapter 3.2.9.3, "OLTR: Results"
The OLTR measurement provides a detailed view for each measurement sequence.
An OLTR limit violation is indicated underneath the graphs:
GUI Reference