1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
433
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:MEAS<i>:OLTR:LIMit:ILOWer
<InitialLower>
Sets initial lower limit for open loop power control step response (3GPP2 C.S0033).
Parameters:
<InitialLower>
Range:
-2 dB to -1 dB
*RST:
-1 dB
Default unit: dB
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.20
Manual operation:
See
3.5.5.3
OLTR Results (Traces)
The following commands return the OLTR trace results
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN?
............................ 433
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP?
.................................. 433
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN?
..............................433
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP?
................................... 433
CALCulate:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN:STATe?
CALCulate:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP:STATe?
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN:STATe?
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP:STATe?
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN:STATe?
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP:STATe?
FETCh:EVDO:MEAS<i>:RPINterval:SEQuence<Sequence>:TRACe:DOWN?
FETCh:EVDO:MEAS<i>:RPINterval:SEQuence<Sequence>:TRACe:UP?
........................... 434
READ:EVDO:MEAS<i>:RPINterval:SEQuence<Sequence>:TRACe:DOWN?
READ:EVDO:MEAS<i>:RPINterval:SEQuence<Sequence>:TRACe:UP?
............................ 435
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN?
FETCh:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP?
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:DOWN?
READ:EVDO:MEAS<i>:OLTR:SEQuence<Sequence>:TRACe:UP?
Returns the values of the OLTR traces. For each sequence, DOWN/UP commands
return the results of the 100 ms interval following the power up/down step.
Suffix:
<Sequence>
.
1..5
The number of the sequence whose trace shall be queried; must
be smaller than the configured number of sequences (see
CONFigure:EVDO:MEAS<i>:OLTR:SEQuence
Return values:
<Reliability>
Command Reference