1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
326
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:MEAS<i>:MEValuation:RESult:POWer
<Enable>
CONFigure:EVDO:MEAS<i>:MEValuation:RESult:TXM
<Enable>
Enables or disables the evaluation of results and shows or hides the views in the multi
evaluation measurement. The mnemonic after "RESult" denotes the view type: Error
Vector Magnitude, Magnitude Error, Phase Error, Adjacent Channel Power, Occupied
Bandwidth, Code Domain Power, Code Domain Error, Channel Power, IQ, Power and
TX Measurement.
Parameters:
<Enable>
OFF | ON
ON:
Evaluate results and show view
OFF:
Do not evaluate results, hide view
*RST:
ON for Spectrum, Power and TX Measurement, all
others OFF
Example:
See
Multi Evaluation: Performing Single-Shot Measurements
Firmware/Software:
V1.0.3.3,
V2.0.10: new command for channel power
V2.0.20: new command for OBW
V3.2.80: new command for IQ
3.5.4.6
Measurement Parameters
The following commands define general settings for the multi evaluation measurement.
CONFigure:EVDO:MEAS<i>:MEValuation:TOUT
.............................................................. 326
CONFigure:EVDO:MEAS<i>:MEValuation:REPetition
........................................................327
CONFigure:EVDO:MEAS<i>:MEValuation:SCONdition
...................................................... 327
CONFigure:EVDO:MEAS<i>:MEValuation:SCOunt:MODulation
..........................................328
CONFigure:EVDO:MEAS<i>:MEValuation:SCOunt:SPECtrum
........................................... 328
CONFigure:EVDO:MEAS<i>:MEValuation:MOEXception
................................................... 329
CONFigure:EVDO:MEAS<i>:MEValuation:ILCMask
.......................................................... 329
CONFigure:EVDO:MEAS<i>:MEValuation:QLCMask
.........................................................329
CONFigure:EVDO:MEAS<i>:MEValuation:RPMode
.......................................................... 329
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:LOWer
...................................... 330
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:UPPer
.......................................330
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:FOFFsets:LOWer
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:FOFFsets:UPPer
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:RBW:LOWer
.............................................332
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:RBW:UPPer
............................................. 332
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:RBW:LOWer
.............................333
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:RBW:UPPer
............................. 333
CONFigure:EVDO:MEAS<i>:MEValuation:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
Command Reference