1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
186
User Manual 1173.9511.02 ─ 12
2.5.8.4
Session Layer
The following parameters configure the session layer. All settings are valid for subtype
0/1 and for subtype 2 signals.
CONFigure:EVDO:SIGN<i>:LAYer:SESSion:ISTimeout
..................................................... 186
CONFigure:EVDO:SIGN<i>:LAYer:SESSion:SNINcluded
...................................................186
CONFigure:EVDO:SIGN<i>:LAYer:SESSion:ISTimeout
<ISTimeout>
Specifies the time after which the AT, if it does not detect any traffic from the
R&S
CMW directed to it, closes the session.
Parameters:
<ISTimeout>
Range:
0 min to 65.535E+3 min
*RST:
3240 min
Example:
See
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:EVDO:SIGN<i>:LAYer:SESSion:SNINcluded
<SNIncluded>
Specifies whether the '"ATISubnetMask" field and the "UATI104" field are included in
the UATI Assignment Message.
Parameters:
<SNIncluded>
OFF | ON
*RST:
OFF
Example:
See
Firmware/Software:
V1.0.15.0
Manual operation:
See
2.5.8.5
Connection Layer
The following parameters configure the connection layer. All settings are valid for sub-
type 0/1 and for subtype 2 signals.
This requires option R&S CMW-KS890.
CONFigure:EVDO:SIGN<i>:LAYer:CONNection:PDTHreshold
........................................... 186
CONFigure:EVDO:SIGN<i>:LAYer:CONNection:RLFoffset
.................................................187
CONFigure:EVDO:SIGN<i>:LAYer:CONNection:ROMessages
........................................... 187
CONFigure:EVDO:SIGN<i>:LAYer:CONNection:PDTHreshold
<PDThreshold>
Defines the pilot power (relative to the total "1xEV-DO Power") below which the AT
starts the pilot supervision timer and eventually announces that it has lost the network
connection.