RF Output (TX)
The following parameters configure the RF output path of the R&S
CMW.
Connector/Converter ← RF Output (TX)
Selects the output path for the generated RF signal, i.e. the output connector and the
TX module to be used.
Depending on your hardware configuration there are dependencies between both
parameters. Select the RF connector first. The "Converter" parameter offers only val-
ues compatible with the selected RF connector.
Remote command:
Depends on the scenario, see
Example:
ROUTe:CDMA:SIGN<i>:SCENario:SCELl
External Attenuation ← RF Output (TX)
Defines the value of an external attenuation (or gain, if the value is negative) in the out-
put path. With an external attenuation of x dB, the power of the generated signal is
increased by x dB. The actual generated levels are equal to the displayed values plus
the external attenuation.
If a correction table for frequency-dependent attenuation is active for the chosen con-
nector, then the table name and a button are displayed. Press the button to display the
table entries.
Remote command:
CONFigure:CDMA:SIGN<i>:RFSettings:EATTenuation
RF Input (RX)
The following parameters configure the RF input path of the R&S
CMW.
Connector/Converter ← RF Input (RX)
Selects the input path for the measured RF signal, i.e. the input connector and the RX
module to be used.
GUI Reference