1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
55
User Manual 1173.9511.02 ─ 12
Fig. 2-17: I/Q settings (standard cell fading: external)
...............................................................................................55
Connector (Out / In)
Selects the output connector. The input connector depends on the output connector
and is displayed for information.
The DIG IQ connectors are located at the rear panel (if an I/Q board is installed).
Remote command:
ROUTe:EVDO:SIGN<i>:SCENario:SCFading[:EXTernal]
Sample Rate (Out / In)
The used sample rate is displayed for information. The value of 100 Msps is fixed.
Configure the connected instrument accordingly (baseband input settings and digital
I/Q output settings).
Remote command:
SENSe:EVDO:SIGN<i>:IQOut:PATH<n>?
Baseband PEP (Out / In)
Indicates the peak envelope power of the baseband signal as dB value relative to full
scale. "Full scale" in this case corresponds to the maximum representable amplitude of
the I/Q samples.
Use the displayed output PEP value to configure the baseband input of the
R&S
AMU200A.
Configure the input PEP so that it matches the baseband output of the connected
instrument.
Remote command:
SENSe:EVDO:SIGN<i>:IQOut:PATH<n>?
CONFigure:EVDO:SIGN<i>:IQIN:PATH<n>
GUI Reference