![Rohde & Schwarz CMW-KE100 User Manual Download Page 855](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807855.webp)
CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
855
User Manual 1173.9511.02 ─ 12
The values described below are returned by
FETCh
commands.
CALCulate
com-
mands return limit check results instead, one value for each result listed below.
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Suffix:
<no>
.
1..200
The segment number must not exceed the total number of seg-
ments measured.
Return values:
<1_Reliability>
. In list mode, a zero reliability indicator indi-
cates that the results in all measured segments are valid. A non-
zero value indicates that an error occurred in at least one of the
measured segments.
<2_SegReliability>
0 | 3 | 4 | 8
The segment reliability indicates whether one of the following
exceptions occurred in this segment:
0:
No error
3:
Signal overflow
4:
Signal low
8:
Synchronization error
If a combination of exceptions occurs, the most severe error is
indicated.
<3_ACPM20> ...
<22_ACPM1>
<23_ACPCarrier>
<24_ACPP1> ...
<43_ACPP20>
Power of adjacent channel.
Range:
-100 dB to 20 dB
Default unit: dB
<44_MSPowWide>
Results of the wideband (8 MHz) power measurement.
Range:
-55 dBm to 42 dBm
Default unit: dBm
<45_MSPowNarrow> Results of the narrowband (1.23 MHz) power measurement.
Range:
-55 dBm to 42dBm
Default unit: dBm
<46_OutOfTolCount> Out of tolerance result, i.e. percentage of measurement intervals
of the statistic count
(
CONFigure:CDMA:MEAS<i>:MEValuation:SCOunt:
MODulation
) exceeding the specified limits.
Range:
0 % to 100 %
Default unit: %
<47_CurStatCount> Number of evaluated valid slots in this segment.
Range:
0 to 1000
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Command Reference
深圳德标仪器
135-1095-0799