1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
426
User Manual 1173.9511.02 ─ 12
Return values:
<Reliability>
See
chapter 3.5.1.6, "Reliability Indicator"
<WideBandPower>
Comma separated list of values, one per active segment.
Range:
-100 dBm to 50 dBm
Default unit: dBm
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
FETCh:EVDO:MEAS<i>:MEValuation:LIST:ACP:OTOLerance?
Returns the out of tolerance percentages in the adjacent channel power measurement
for all active list mode segments.
Return values:
<Reliability>
chapter 3.5.1.6, "Reliability Indicator"
<OutOfTolCount>
Comma separated list of values, one per active segment.
Range:
0 % to 100 %
Default unit: %
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
FETCh:EVDO:MEAS<i>:MEValuation:LIST:ACP:STCount?
Returns the statistic count of the spectrum measurement for all active list mode seg-
ments.
Return values:
<Reliability>
chapter 3.5.1.6, "Reliability Indicator"
<StatisticCount>
Number of evaluated valid slots.
Comma separated list of values, one per active segment.
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
Command Reference