1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
300
User Manual 1173.9511.02 ─ 12
Multi Evaluation Measurement Commands
.......................................................... 314
.......................................................................... 427
.......................................................................435
3.5.1 Conventions and General Information
The following sections describe the most important conventions and general informa-
tion concerning the command reference.
3.5.1.1
MEAS<i>
MEAS<i> is used as abbreviation of "MEASurement<instance>". For better readability
only the abbreviated form (which is also accepted by the instrument) is given in the
command reference.
The <instance> is relevant for instruments supporting several instances of the same
firmware application. It can be omitted if the instrument supports only one instance, or
to address the first instance.
See also: "Firmware Applications" in the R&S
CMW user manual, chapter "Remote
Control"
3.5.1.2
FETCh, READ and CALCulate Commands
All commands are used to retrieve measurement results:
●
FETCh...
returns the results of the current measurement cycle (single-shot mea-
surement) after they are valid.
FETCh...
must be used after the measurement
has been started (
INITiate...
, measurement states RUN or RDY).
●
READ...
starts a new single-shot measurement and returns the results.
●
CALCulate...
returns one limit check result per
FETCh
result:
–
OK
: The
FETCh
result is located within the limits or no limit has been defined/
enabled for this result.
–
ULEU
(User limit exceeded upper): An upper limit is violated. The
FETCh
result
is located above the limit.
–
ULEL
(User limit exceeded lower): A lower limit is violated. The
FETCh
result is
located below the limit.
See also: "Retrieving Measurement Results" in the R&S
CMW user manual, chapter
"Remote Control"
3.5.1.3
Current and Statistical Results
The R&S
CMW repeats measurements according to the selected statistic count and
repetition mode. Consecutive measurement values are stored and used to calculate
statistical results, e.g. average, minimum, maximum and standard deviation.
Command Reference