![Rohde & Schwarz CMW-KE100 User Manual Download Page 255](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807255.webp)
1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
255
User Manual 1173.9511.02 ─ 12
Fig. 3-13: 1xEV-DO multi evaluation: spectrum (tabular view)
The description of statistical measurement results is covered in
Evaluation Detailed Views: TX Measurement (Scalar) and Power"
For additional information refer to
3.2.8.5
Multi Evaluation Detailed Views: Channel Power
The Channel Power results are displayed in bar graphs of the reverse physical chan-
nels of both the I and Q signal. The power values are measured in dB relative to the
selected reference power: see
on page 277 for details.
Separate graphs are available for the I-branch and the Q-branch of the signal for sub-
type 0/1. For subtype 2 and 3 signals, the I/Q setting table is not shown. Instead the
measured signals are shown as bar graphs in one diagram and the in-phase and quad-
rature components of Walsh codes are displayed in a separated table below.
Bar graphs in red indicate they exceed the limit set in the Configuration Menu. The limit
lines provide a quick reference point to view channels exceeding set limits. The limit
line level is set in the Configuration Menu. Bars for active code channels are colored.
Bars for inactive code channels are gray.
Below the bar graphs statistical information is displayed.
General Description