1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
314
User Manual 1173.9511.02 ─ 12
TRIGger:EVDO:MEAS<i>:MEValuation:MGAP
<MinTriggerGap>
Sets a minimum time during which the IF signal must be below the trigger threshold
before the trigger is armed so that an IF power trigger event can be generated.
Parameters:
<MinTriggerGap>
Range:
0 s to 0.01 s
Increment: 0.00000001 (0.01 µs)
*RST:
0 s
Default unit: s
Example:
See
Configuring the Trigger System
Firmware/Software:
V2.0.10
Manual operation:
See
TRIGger:EVDO:MEAS<i>:MEValuation:EOFFset
<EvalOffset>
Defines a delay time for the measurement relative to the "IF Power" or external trigger
events (see
TRIGger:EVDO:MEAS<i>:MEValuation:SOURce
entered as an integer number of slots. Each slot has a duration of 1.25 ms.
Parameters:
<EvalOffset>
Range:
0 to 64
*RST:
0
Default unit: slot
Example:
See
Configuring the Trigger System
Firmware/Software:
V2.0.20
3.5.4 Multi Evaluation Measurement Commands
The commands for the 1xEV-DO "Multi Evaluation" measurement are divided into the
groups listed below. The general measurement settings also affect the measurement,
see
.
Measurement Control and States
......................................................................... 315
.............................................................................................317
.............................................................................................319
..................................................................................324
.....................................................................................326
................................................................................................334
........................................................................................... 338
...........................................................................................346
Magnitude Error Results (Traces)
.........................................................................346
................................................................................347
IQ Constallation Results (Traces)
......................................................................... 347
........................................................................................... 348