1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
315
User Manual 1173.9511.02 ─ 12
........................................................................................... 353
................................................................................................ 360
Modulation Results (Single Values)
...................................................................... 369
.................................................................................................371
List Mode Results: Result Groups, Single Segment
............................................. 377
List Mode Results: Result Groups, All Segments
................................................. 390
List Mode Results: Single Results, All Segments
................................................. 403
3.5.4.1
Measurement Control and States
The following commands control the measurement and return the current measure-
ment state.
INITiate:EVDO:MEAS<i>:MEValuation
............................................................................. 315
................................................................................315
ABORt:EVDO:MEAS<i>:MEValuation
.............................................................................. 315
FETCh:EVDO:MEAS<i>:MEValuation:STATe?
................................................................. 316
FETCh:EVDO:MEAS<i>:MEValuation:STATe:ALL?
...........................................................316
INITiate:EVDO:MEAS<i>:MEValuation
STOP:EVDO:MEAS<i>:MEValuation
ABORt:EVDO:MEAS<i>:MEValuation
Starts, stops, or aborts the measurement:
●
INITiate...
starts or restarts the measurement; the R&S
CMW enters the
"RUN" state.
●
STOP...
causes a running measurement to stop after the current evaluation
period is terminated and valid results are available; the R&S
CMW enters the
"RDY" state.
●
ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Example:
See
Multi Evaluation: Single-Shot and Continuous Measure-
Usage:
Event
Firmware/Software:
V1.0.3.3
Manual operation:
See
Command Reference