![Rohde & Schwarz CMW-KE100 User Manual Download Page 317](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807317.webp)
1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
317
User Manual 1173.9511.02 ─ 12
<SyncState>
PEND | ADJ | INV
PEND:
waiting for resource allocation, adjustment, hardware
switching ("pending")
ADJ:
all necessary adjustments finished, measurement running
("adjusted")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
<ResourceState>
QUE | ACT | INV
QUE:
measurement without resources, no results available
("queued")
ACT:
resources allocated, acquisition of results in progress but
not complete ("active")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
Example:
See
Multi Evaluation: Single-Shot and Continuous Measure-
Usage:
Query only
Firmware/Software:
V1.0.3.3
Manual operation:
See
3.5.4.2
Carrier Configuration
The following commands configure the reverse link carriers to be measured.
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:SETTing
.............................................317
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:ENABle
.............................................. 318
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FOFFset
.............................................318
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FREQuency
........................................319
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:WBFilter
............................................. 319
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:SETTing
<SetCarrier>
Selects a carrier for the following carrier settings:
●
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:ENABle
●
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FOFFset
●
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FREQuency
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use:
●
CONFigure:EVDO:SIGN<i>:CARRier:SETTing
●
CONFigure:EVDO:SIGN<i>:PILot:SETTing
Parameters:
<SetCarrier>
Range:
0 to 2
*RST:
0
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Command Reference
135-1095-0799