1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
318
User Manual 1173.9511.02 ─ 12
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:ENABle
<CEnable>
Defines whether a carrier shall be measured (ON) or not (OFF).
The related carrier has to be pre-set using the
command.
All carriers can be queried, but carrier 0 cannot be set (fix set to ON).
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use:
●
CONFigure:EVDO:SIGN<i>:NETWork:PILot:CSECtor
●
CONFigure:EVDO:SIGN<i>:NETWork:PILot:AN:ACTive
●
CONFigure:EVDO:SIGN<i>:NETWork:PILot:AT:ASSigned
Parameters:
<CEnable>
OFF | ON
*RST:
carrier 0 = ON, carrier 1/2 = OFF
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FOFFset
<CFOffset>
Gets/sets the frequency offset of a selected carrier relative to carrier 0.
The related carrier has to be pre-set using the
command.
All carriers can be queried, but carrier 0 cannot be set.
This command is relevant only for standalone mode. While the combined signal path
scenario is active, the command for carrier frequency offset is not used.
Parameters:
<CFOffset>
The offset relative to carrier 0. The maximum distance between
carriers is restricted to 8 MHz.
Range:
- 8E+6 Hz to + 8E+6 Hz
*RST:
0 Hz
Default unit: Hz
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
"Offset to RF / Carrier Freq."
Command Reference