2.3.7 System Settings
The parameters in this section provide general signal settings, system settings and
configure the RF input and output path.
Fig. 2-22: System settings
...............................................................................................65
..........................................................................................67
...........................................................................................67
2.3.7.1
RF Output (TX)
Fig. 2-23: 1xEV-DO RF output (TX) settings
Routing
Selects the output path for the generated RF signal, i.e. the output connector and the
TX module to be used.
Depending on your hardware configuration there are dependencies between both
parameters. Select the RF connector first. The "Converter" parameter offers only val-
ues compatible with the selected RF connector.
Remote command:
Depends on the scenario, see
chapter 2.5.6.1, "Signal Routing"
Example:
ROUTe:EVDO:SIGN<i>:SCENario:SCELl
External Attenuation
Defines the value of an external attenuation (or gain, if the value is negative) in the out-
put path. With an external attenuation of x dB, the power of the generated signal is
increased by x dB. The actual generated levels are equal to the displayed values plus
the external attenuation.