1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
211
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:SIGN<i>:RXQuality:PER:TOUT
<Timeout>
CONFigure:EVDO:SIGN<i>:RXQuality:THRoughput:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Firmware/Software:
V3.0.10
Manual operation:
See
2.5.11.3
Measurement States
The following commands return the current measurement states.
.............................................................................. 211
FETCh:EVDO:SIGN<i>:THRoughput:STATe?
...................................................................211
FETCh:EVDO:SIGN<i>:PER:STATe:ALL?
........................................................................212
FETCh:EVDO:SIGN<i>:THRoughput:STATe:ALL?
............................................................212
FETCh:EVDO:SIGN<i>:PER:STATe?
FETCh:EVDO:SIGN<i>:THRoughput:STATe?
Queries the main measurement state. Use
FETCh:...:STATe:ALL?
to query the
measurement state including the substates. Use
INITiate..., STOP...,
ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Return values:
<State>
OFF | RDY | RUN
OFF:
measurement switched off, no resources allocated, no
results available (when entered after
ABORt...
)
RDY:
measurement has been terminated, valid results may be
available
RUN:
measurement running (after
INITiate..., READ...
),
synchronization pending or adjusted, resources active or queued
Command Reference
135-1095-0799