1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
172
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:SIGN<i>:APPLication:DSIGnaling?
Queries the stream and the state of the Default Signaling Application. The response is
fixed: The Default Signaling Application is always enabled and assigned to stream 0.
Return values:
<Stream>
Range:
0
Usage:
Query only
Firmware/Software:
V1.0.15.0
CONFigure:EVDO:SIGN<i>:APPLication:MODE
<Mode>
Selects the application to be instantiated.
Parameters:
<Mode>
FWD | REV | FAR | PACKet
FWD
: forward test application
REV
: reverse test application
FAR
: forward and reverse test application
PACKet
: packet application
*RST:
FWD
Example:
See
Configuring the Test Application
Firmware/Software:
V1.0.15.0
V3.0.10: modified command syntax (token TEST removed),
PACKet mode added
Options:
PACKet mode requires R&S CMW-KA100
Manual operation:
See
CONFigure:EVDO:SIGN<i>:APPLication
<Stream>
Queries the stream occupied by the configured session application (see
Setting this value will have no effect as the selected stream is a result of the session
negotiation.
Parameters:
<Stream>
Range:
1 to 3
Example:
See
Configuring the Test Application
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:EVDO:SIGN<i>:LAYer:APPLication:PACKet:MODE
<PrefMode>
Sets the preferred standard to be signaled during the connection setup.
Command Reference