![Rohde & Schwarz CMW-KE100 User Manual Download Page 771](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807771.webp)
CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
771
User Manual 1173.9511.02 ─ 12
CONFigure:CDMA:MEAS:MEValuation:LIMit:EVMagnitude 20, 40
CONFigure:CDMA:MEAS:MEValuation:LIMit:PERRor 20, OFF
CONFigure:CDMA:MEAS:MEValuation:LIMit:IQOFfset -20
CONFigure:CDMA:MEAS:MEValuation:LIMit:IQIMbalance ON
CONFigure:CDMA:MEAS:MEValuation:LIMit:CFERror 150
CONFigure:CDMA:MEAS:MEValuation:LIMit:TTERror 1
CONFigure:CDMA:MEAS:MEValuation:LIMit:WFQuality 0.95
CONFigure:CDMA:MEAS:MEValuation:LIMit:POWer -20, -10
// *************************************************************************
// Define limits for code domain power, code domain error,
// channel power, channel phase offset and for the channel time
// offset. Activate the IQ leakage check.
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:LIMit:CDP -20
CONFigure:CDMA:MEAS:MEValuation:LIMit:CDE -20
CONFigure:CDMA:MEAS:MEValuation:LIMit:CP -20
CONFigure:CDMA:MEAS:MEValuation:LIMit:CPO 100
CONFigure:CDMA:MEAS:MEValuation:LIMit:CTO 20
CONFigure:CDMA:MEAS:MEValuation:IQLCheck ON
// *************************************************************************
// Define ACP limits
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:LIMit:ACP:RELative 5,0,-5,-10,-15,-20
CONFigure:CDMA:MEAS:MEValuation:LIMit:ACP:ABSolute 5,0,-5,-10, OFF, OFF
// *************************************************************************
// Define OBW limit 2.96 MHz
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:LIMit:OBW 2.96
5.4.1.5
Multi Evaluation: Performing Single-Shot Measurements
// *************************************************************************
// Enable all measurements
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:RESult:EVMagnitude ON
CONFigure:CDMA:MEAS:MEValuation:RESult:MERRor ON
CONFigure:CDMA:MEAS:MEValuation:RESult:PERRor ON
CONFigure:CDMA:MEAS:MEValuation:RESult:IQ ON
CONFigure:CDMA:MEAS:MEValuation:RESult:ACP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CDP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CDE ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CPO ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CTO ON
CONFigure:CDMA:MEAS:MEValuation:RESult:POWer ON
CONFigure:CDMA:MEAS:MEValuation:RESult:MODQuality ON
Programming
深圳德标仪器
135-1095-0799