CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
697
User Manual 1173.9511.02 ─ 12
CONFigure:CDMA:SIGN<i>:RXQuality:FERFch:TOUT
<Timeout>
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Example:
See
Firmware/Software:
V3.0.20
Manual operation:
See
CONFigure:CDMA:SIGN<i>:RXQuality:FERFch:REPetition
<Repetition>
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:REPetition
<Repetition>
Specifies the repetition mode of the measurement. The repetition mode specifies
whether the measurement is stopped after a single-shot or repeated continuously. Use
CONFigure:CDMA:SIGN<i>:RXQuality:FERSch:FRAMes
to determine the num-
ber of test frames per single shot.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot
: Single-shot measurement
CONTinuous
: Continuous measurement
*RST:
SINGleshot
Example:
See
Firmware/Software:
V1.0.15.0
V3.0.10 Command for FCH renamed (token
TDATa
added)
V3.0.20 Command for FCH renamed (token
TDATa
replaced by
FERFch
)
V3.0.20 SCH
Manual operation:
See
Command Reference