1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
127
User Manual 1173.9511.02 ─ 12
RF path selection values:
●
RX module:
RX1 | RX2 | RX3 | RX4
●
TX module:
TX1 | TX2 | TX3 | TX4
●
RX connector:
RF1C | RF2C | RF3C | RF4C | RFAC | RFBC
RF 1 COM to RF 4 COM plus virtual connector names
●
TX connector:
RF1C | RF2C | RF3C | RF4C | RFAC | RFBC
RF1O | RF3O | RFAO
RF 1 COM to RF 4 COM plus virtual connector names
RF 1 OUT and RF 3 OUT plus virtual connector name
I/Q path selection values:
●
I/Q output connector:
IQ2O | IQ4O | IQ6O | IQ8O
DIG IQ OUT 2 to 8
2.5.1.4
Keywords
Selected keywords used in the command description are described in the following.
●
Command usage
If the usage is not explicitly stated, the command allows you to set parameters and
query parameters. Otherwise the command usage is stated as follows:
–
"Setting only": command can only be used to set parameters
–
"Query only": command can only be used to query parameters
–
"Event": command initiates an event
●
Parameter usage
The parameter usage is indicated by the keyword preceding the parameter(s):
–
"Parameters" are sent with a setting or query command and are returned as
the result of a query
–
"Setting parameters" are only sent with a setting command
–
"Query parameters" are only sent with a query command (to refine the query)
–
"Return values" are only returned as the result of a query
●
Firmware/Software:
Indicates the lowest software version supporting the command. Command
enhancements in later software versions are also indicated.
2.5.1.5
Reliability Indicator
The reliability indicator has one of the following values:
●
0 (OK)
:
Measurement values available, no error detected.
●
1 (Measurement Timeout)
:
Command Reference