1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
345
User Manual 1173.9511.02 ─ 12
Parameters:
<OBWLimitCheck>
OFF | ON
Disable/enable OBW limit checks.
*RST:
ON
Example:
See
Firmware/Software:
2.1.20
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:OBW:CHECk:USED?
Returns the currently used OBW limit set. This is ultimately determined by the currently
selected band class: limit set A is dedicated to band class 0, limit set B is used for all
other band classes.
Return values:
<OBWUsedLimitSet> SETA | SETB
The currently used limit set.
Example:
See
Usage:
Query only
Firmware/Software:
V.2.1.20
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:OBW:SETA
<OBWLimit_1>,<OBWLimit2>,<OBWLimit3>
Gets/sets the OBW limits for Limit Set A. This limit set is dedicated to Band Class 0.
Parameters:
<OBWLimit_1i>,
<OBWLimit_2a>,
<OBWLimit_3a>
Range:
0 MHz to 16 MHz
*RST:
1.48 MHz
The OBW limits for 1 isolated carrier, 2 adjacent carriers and 3
adjacent carriers. In single-carrier scenarios, OBWLimit_1i is the
one and only applicable OBW limit.
Example:
See
Firmware/Software:
V2.1.20
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:LIMit:OBW:SETB
<OBWLimit_1i>,<OBWLimit2a>,<OBWLimit3a>
Gets/sets the OBW limits for Limit Set B. This limit set is used if the currently selected
Band Class is different from 0.
Command Reference