1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
331
User Manual 1173.9511.02 ─ 12
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:FOFFsets:LOWer
<FrequencyOffset>(20)
Defines the negative (lower) frequency offsets 19 to 0, to be used for extended ACP
measurements of the selected carrier (see
). The offsets are defined relative to the analyzer
frequency.
Up to 20 offsets can be defined and enabled. The offset index 19 to 0 corresponds to
the index used in manual control.
Parameters:
<FrequencyOffset>
Range:
-4 MHz to 0 MHz
*RST:
Offset 19 to 0 [MHz]: depends on the carrier selec-
tion
Default unit: Hz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V3.2.20
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:EXTended:FOFFsets:UPPer
<FrequencyOffset>(20)
Defines the positive (upper) frequency offsets 0 to 19, to be used for extended ACP
measurements of the selected carrier (see
). The offsets are defined relative to the analyzer
frequency.
Up to 20 offsets can be defined and enabled. The offset index 0 to 19 corresponds to
the index used in manual control.
Parameters:
<FrequencyOffset>
Range:
0 MHz to 4 MHz
*RST:
Offset 0 to 19 [MHz]: depends on the carrier selec-
tion
Default unit: Hz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V3.2.20
Manual operation:
See
Command Reference