1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
330
User Manual 1173.9511.02 ─ 12
Parameters:
<RefPowerMode>
ATPower | PPOWer
ATPower
: total channel power
PPOWer
: pilot power
*RST:
ATP
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
2.1.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:LOWer
<FreqOffset9>, ..., <FreqOffset0>
Defines the negative (lower) frequency offsets to be used for ACP measurements of
the selected carrier (see
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:
). The offsets are defined relative to the analyzer frequency.
Up to 10 offsets can be defined and enabled. The offset index 0 to 9 corresponds to
the index used in manual control.
Parameters:
<FreqOffset9> ...
<FreqOffset0>
Range:
-4 MHz to 0 MHz
*RST:
Offset 9 to 0 [MHz]: -4.0, -3.5, -3.0, -2.5, -2.0, -1.6,
-1.2, -0.9, -0.885, -0.87
Default unit: MHz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:UPPer
<FreqOffset0>, ..., <FreqOffset9>
Defines the positive (upper) frequency offsets to be used for ACP measurements of the
selected carrier (see
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:
). The offsets are defined relative to the analyzer frequency.
Up to 10 offsets can be defined and enabled. The offset index 0 to 9 corresponds to
the index used in manual control.
Parameters:
<FreqOffset0> ...
<FreqOffset9>
Range:
0 MHz to 4 MHz
*RST:
Offset 0 to 9 [MHz]: 0.87, 0.885, 0.9, 1.2, 1.6, 2.0,
2.5, 3.0, 3.5, 4.0
Default unit: MHz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Command Reference