1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
167
User Manual 1173.9511.02 ─ 12
CONFigure:EVDO:SIGN<i>:RPControl:REPetition
<Repetition>
Specifies the repetition mode of the pattern execution.
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot
: the pattern execution is stopped after a single-shot
CONTinuous
: the pattern execution is repeated continuously
and stopped by the
CONFigure:EVDO:SIGN<i>:RPControl:
Example:
See
Configuring Reverse Power Control Setup
Firmware/Software:
V3.2.80
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:BITS
<SegmentBits>
Sets the user specific power control bits.
Suffix:
<no>
.
1..4
Segment number
Parameters:
<SegmentBits>
DOWN | UP | ALTernating
All 0, all 1 or alternating
Example:
See
Configuring Reverse Power Control Setup
Firmware/Software:
V3.2.80
Options:
R&S CMW-K890
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RPControl:SEGMent<no>:LENGth
<SegmentLength>
Sets the length of the segment of the user specific power control bits.
Suffix:
<no>
.
1..4
Segment number
Parameters:
<SegmentLength>
Segment length
Range:
0 bits to 128 bits
Default unit: bit
Example:
See
Configuring Reverse Power Control Setup
Firmware/Software:
V3.2.80
Options:
R&S CMW-K890
Command Reference