![R&S RTO6 Скачать руководство пользователя страница 735](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646735.webp)
Protocol analysis
R&S
®
RTO6
735
User Manual 1801.6687.02 ─ 05
"Condition"
Sets the comparison condition to a specific value or a range.
" Source
Address (Min)"
Specifies the value or sets the start value of a range.
"Source
Adress (Max)"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:ETHernet:FRAMe:SCONdition
TRIGger<m>:ETHernet:FRAMe:SMAX
TRIGger<m>:ETHernet:FRAMe:SMIN
Length/ Type
Defines the length/type field value.
"Condition"
Sets the operator to trigger on a specific address pattern ("Equal" or
"Not equal") or an address range.
"Length/Type
(Min)"
Defines the length/type pattern for all operators that require one pat-
tern.
"Length/Typ
(Max)"
Defines the length/type pattern that is required to specify a range with
conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:ETHernet:FRAMe:TCONdition
TRIGger<m>:ETHernet:FRAMe:TMAX
TRIGger<m>:ETHernet:FRAMe:TMIN
Frame check
Defines the frame check sequence value.
"Condition"
Sets the comparison condition to a specific value or a range.
"Frame check
(Min)"
Specifies the value or sets the start value of a range.
"Frame check
(Max)"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:ETHernet:FRAMe:CCONdition
TRIGger<m>:ETHernet:FRAMe:CMAX
TRIGger<m>:ETHernet:FRAMe:CMIN
Error
Triggers on enabled errors.
"Preamble
error"
Triggers on a frame with invalid preamble.
"Length error"
Triggers on an incorrect length of the sequence - when additional or
missing bits are detected and the sequence of bits is not as expected.
"CRC error"
Triggers on a frame that has a mismatch of the cyclic redundancy
check (CRC) value between the transmitting and receiving device.
Ethernet 10BASE-T and 100BASE-TX (option R&S