![R&S RTO6 Скачать руководство пользователя страница 1580](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461580.webp)
Remote control commands
R&S
®
RTO6
1580
User Manual 1801.6687.02 ─ 05
24.13.3
Spectrum analysis (option R&S
RTO6-K37)
In all
CALC:MATH<m>:FFT
commands, the suffix <m> selects the math waveform.
In all
CALC:MATH<m>:FFT:SPEC:TIM
commands, the suffix <m> selects the timeline.
CALCulate:MATH<m>:FFT:SPECtrogram:CMODe
.......................................................... 1580
CALCulate:MATH<m>:FFT:USEColtab
...........................................................................1580
CALCulate:MATH<m>:FFT:SPECtrogram:STATe
.............................................................1581
CALCulate:MATH<m>:FFT:SPECtrogram:TIMeline<n>:POSition
...................................... 1581
CALCulate:MATH<m>:FFT:SPECtrogram:TIMeline<n>:STATe
..........................................1581
CALCulate:MATH<m>:FFT:SPECtrogram:CMODe
<ColorTableMode>
Selects the color table mode for the frequency analysis display.
Suffix:
<m>
.
1..8
Parameters:
<ColorTableMode>
INCI | AMPL
INCI
("Incidence") The display color is set depending on the fre-
quency of occurrence of a value.
AMPL
("Amplitude") In the spectrogram and the frequency analysis dis-
play, the color is used to indicate the magnitude of the FFT sig-
nal. The higher the amplitude the higher the color in the
assigned color table.
*RST:
INCI
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:USEColtab
<UseColorTable>
If enabled, the selected waveform is displayed according to its assigned color table.
For information on the available color tables, see
If this option is disabled, the preset color of the selected channel source is displayed,
and the intensity of the specific signal color varies according to the cumulative occur-
rence of the values.
Suffix:
<m>
.
1..8
Parameters:
<UseColorTable>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Spectrum analysis