![R&S RTO6 Скачать руководство пользователя страница 387](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646387.webp)
Spectrum analysis
R&S
®
RTO6
387
User Manual 1801.6687.02 ─ 05
Coupled to M1...M8
Copies the current FFT settings of the selected math waveform (M1...M8) to the other
selected math waveforms, and couples those waveforms. Thus, if any FFT setting for
any of the coupled spectrums are changed, they are changed for all coupled spec-
trums.
Two different sets of spectrums can be coupled at the same time, for instance "M1" can
be coupled to "M2", while "M3" is coupled to "M4".
Note that the formula of the coupled math waveforms is not changed. If necessary, you
must select an FFT function for the math waveform manually before the FFT settings
of the coupled waveform are applied. See
Chapter 7.3.1, "Displaying math wave-
Remote command:
CALCulate:MATH<m>:FFT:COUPled:WITH<1..8>
9.1.4
FFT scale settings
In this section, you can find functions to set the vertical parameters of the FFT math
waveform.
FFT analysis