Usage:
Query only
BUS<m>:MDIO:WORD<n>:PHYS?
Returns the 5-bit address field content (PHYAD/PRTAD) in the present acquisition of
the selected word and the selected serial bus.
Suffix:
<m>
.
1..4
<n>
*
Return values:
<PhyAd>
Range:
0 to 32
Increment: 1
*RST:
0
Usage:
Query only
BUS<m>:MDIO:WORD<n>:REGI?
Returns the 5-bit register or device address field content (REGAD/DEVAD) in the pres-
ent acquisition of the selected word and the selected serial bus.
Suffix:
<m>
.
1..4
<n>
*
Return values:
<RegAd>
Range:
0 to 32
Increment: 1
*RST:
0
Usage:
Query only
BUS<m>:MDIO:WORD<n>:ST?
Returns the Start Code (= start of frame code) in the present acquisition of the selected
word and the selected serial bus.
Suffix:
<m>
.
1..4
<n>
*
Return values:
<StartCode>
Range:
0 to 3
Increment: 1
*RST:
0
The parameter value "0" represents Clause 45, and "1" stands
for Clause 22.
The values "2" and "3" do not correspond with any legal parame-
ters according to the standard, but they can be searched for.
Protocols