Remote control commands
R&S
®
RTO6
2563
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n>
2..n-1
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<IsEnabled>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
DEEMbedding<m>:COMPonent<n>:TYPE
<Type>
Sets the type of the indicated component. The first component is always DUT, the last
component is PROBe or INPut.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n>
1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<Type>
PROBe | FIXTure | CABLe | DUT | ADAPter | INPut | OTHer |
RTZA10 | RTZA16 | RTZA17 | PP | PC
RTZA10 | RTZA16 | RTZA17
SMA adapters and cable offered by Rohde
&
Schwarz
PP | PC
Proven probe, proven cable
*RST:
CABLe
Usage:
Asynchronous command
DEEMbedding<m>:COMPonent<n>:NAME
<Name>
Sets a name for the indicated component that helps you identifying it.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n>
1..n
Index of the component in the measurement setup. The DUT
has index 1, the probe has index n.
Parameters:
<Name>
String parameter, name of the component
Deembedding (option R&S
RTO6-K121)