Remote control commands
R&S
®
RTO6
1682
User Manual 1801.6687.02 ─ 05
Parameters:
<UseAddress>
ON | OFF
*RST:
OFF
TRIGger<m>:I2C:ADOR<n>:ADRType
<AddressType>
Sets the address type for the indicated ADOR address in the "address OR" trigger con-
dition.
Suffix:
<n>
.
1..4
Index of the address in an "address OR" condition (OR slot)
Parameters:
<AddressType>
BIT7 | BIT7_RW | BIT10
*RST:
BIT7
TRIGger<m>:I2C:ADOR<n>[:VALue]
<Address>
Defines the address pattern of the indicated ADOR address in the "address OR" trigger
condition.
Suffix:
<n>
.
1..4
Index of the address in an "address OR" condition (OR slot)
Parameters:
<Address>
Numeric or string pattern, see
TRIGger<m>:I2C:ADOR<n>:RWBit
<RWBit>
Defines the R/W bit of the indicated ADOR address in the "address OR" trigger condi-
tion.
Suffix:
<n>
.
1..4
Index of the address in an "address OR" condition (OR slot)
Parameters:
<RWBit>
UNDefined | READ | WRITe | EITHer
UNDefined
Return value only
*RST:
EITHer