Remote control commands
R&S
®
RTO6
2059
User Manual 1801.6687.02 ─ 05
BUS<m>:RFFE:GFWidth
<GlitchFilterWdt>
Sets the maximum glitch width to be ignored.
Suffix:
<m>
.
1..4
Parameters:
<GlitchFilterWdt>
Range:
10E-12 to 10E-9
Increment: 10E-12
*RST:
1E-9
Default unit: s
BUS<m>:RFFE:MINGap:SELect
<MinGapSelect>
If the minimum gap is enabled, the instrument detects the specified gap.
Set the minimum gap time using
Suffix:
<m>
.
1..4
Parameters:
<MinGapSelect>
ON | OFF
*RST:
OFF
BUS<m>:RFFE:MINGap:TIME
<MinGapTime>
Sets the minimum idle time between the Bus Park Cycle (BP) and Sequence Start
Condition (SSC).
The setting is only relevant if
is enabled.
Suffix:
<m>
.
1..4
Parameters:
<MinGapTime>
Range:
1E-9 to 10E-6
Increment: 1E-9
*RST:
10E-9
Default unit: s
BUS<m>:RFFE:RDMD
<ReadMode>
Selects, if the standard or synchronous read (
SREAD
) mode is used.
Suffix:
<m>
.
1..4
Parameters:
<ReadMode>
STRD | SREAD
*RST:
STRD
Protocols