Remote control commands
R&S
®
RTO6
1968
User Manual 1801.6687.02 ─ 05
TRIGger<m>:MILStd:MINGap:BITS
<MinGapTime>
Sets the value for the minimum gap to be triggered on if
is set to
ERRor
.
Parameters:
<MinGapTime>
Range:
2E-6 to 262E-6
Increment: 100E-9
*RST:
4E-6
Default unit: s
TRIGger<m>:MILStd:MINGap:SELect
<MinGapSelect>
Enables triggering when the minimum gap is out of range if
is set to
ERRor
. You can set the minimum gap with:
.
Parameters:
<MinGapSelect>
ON | OFF
*RST:
ON
TRIGger<m>:MILStd:STATus:BCReceived
<BroadcastCmd>
Triggers on the state of the broadcast command received bit of the status word if
is set to
STATword
.
Parameters:
<BroadcastCmd>
ONE | ZERO | DC
*RST:
DC
TRIGger<m>:MILStd:STATus:BUSY
<BusyFlag>
Triggers on the state of the busy bit of the status word if
is set to
STATword
.
Parameters:
<BusyFlag>
ONE | ZERO | DC
*RST:
DC
TRIGger<m>:MILStd:STATus:DBCaccept
<DynBusControl>
Triggers on the state of the dynamic bus control accept bit of the status word if
is set to
STATword
.
Parameters:
<DynBusControl>
ONE | ZERO | DC
*RST:
DC
Protocols