Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1156
User Manual 1801.6687.02 ─ 05
19.4.2.1
Dynamic on resistance results
The results of "Dynamic on Resistance" measurements are provided in two ways:
●
The diagram shows the graphical presentation of:
–
the voltage waveform
–
the current waveform
●
The result box displays the dynamic on resistance value.
To measure and display the dynamic on resistance, the instrument uses the following
measurements and waveforms:
●
"MG 5" to measure the amplitude of the voltage
●
"MG 6" to measure the amplitude of the voltage
●
"MG 7" to measure the amplitude of the current
●
"MG 8" to measure the amplitude of the current
●
"Cu 1 " Cursor 1 to measure gate ["t
0
" ,"t
1
"]
●
"Cu 2 " Cursor 2 to measure gate ["t
2
" ,"t
3
"]
The used resources are listed in the "Details" tab. See also:
The dynamic on resistance displayed as the result is defined as:
The points "t
0
" , "t
1
", "t
2
" and "t
3
" are defined by the cursor lines displayed in the result
diagram of the measurement. You can move the cursor lines to define another area of
interest.
The following remote commands are used for handling the measurement results:
●
POWer:DONRes:RESult:RESistance?
●
●
●
19.4.2.2
Configuring dynamic on resistance
For details of the configuration settings, see
Chapter 19.4.2.3, "Dynamic on resistance
Control loop