Acquisition and waveform setup
R&S
®
RTO6
158
User Manual 1801.6687.02 ─ 05
Setup parameters of modular probes
....................................................................158
Deembedding for modular probes
.........................................................................159
....................................... 161
5.6.5.1
Setup parameters of modular probes
The basic setup parameters of all modular probes are the measurement mode and the
offset settings.
Access: "Menu" > "Vertical" > "Probes"
Make sure to adjust also the deembedding settings under "Probe Deembedding" (see
Chapter 5.6.5.2, "Deembedding for modular probes"
Probe mode
Sets the measurement mode of modular probes.
The modular probes of the R&S
RT
‑
ZM family have a multi-mode function. You can
switch between single-ended, differential and common mode measurements without
reconnecting or resoldering the probe. You can set the probe mode in the dialog box,
and you can assign the probe mode setting to the micro button.
If you use the R&S
RT
‑
ZMA30 browser module, only DM measurements are possible
because this module has no ground connector.
The measurement modes are:
Probes