Remote control commands
R&S
®
RTO6
1417
User Manual 1801.6687.02 ─ 05
TRIGger<m>:TV:CUSTom:SDURation
<SyncPulseDurat>
Sets the width of the sync pulse. Only relevant if
CUSTom
.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SyncPulseDurat>
Range:
100E-9 to 100E-6
Increment: 100E-9
*RST:
4.7E-6
Default unit: s
Usage:
Asynchronous command
24.9.15
Trigger qualification
The A-event can have its own trigger qualification. Qualification is not available for B-
and R-events (event suffix m = 2 | 3) and some trigger types.
A suffix defines the trigger type to which the qualification belongs.
Table 24-3: Trigger type suffixes
Suffix
Trigger type
1
EDGE
2
GLITch
3
WIDTh
4
RUNT
5
WINDow
6
TIMeout
7
INTerval
8
Qualification is not supported (SLEWrate)
9
Qualification is not supported (SETHold)
10
STATe
11
PATTern
12
qualification is not supported (ANEDge, ext. trigger input)
13
Currently not used
14
Currently not used
15
Qualification is not supported (SERPattern)
16
Qualification is not supported (NFC)
Trigger