Getting started
R&S
®
RTO6
38
User Manual 1801.6687.02 ─ 05
[Gen]
Opens the "Waveform Generator" dialog, if option R&S
RTO6-B6 is installed.
The waveform generator can generate various function and arbitrary waveforms,
sweeps, and parallel patterns. For detailed specifications, refer to the data sheet.
[Logic]
Opens the "Logic" dialog to configure and enable the logic buses L1 to L4. A logic bus
(or parallel bus) has up to 16 logic (digital) channels. If logic buses are active, pressing
the key toggles these buses.
The key lights up if at least one logic bus is enabled.
[Bus]
Opens and closes the "Protocol" tab in the "Apps" dialog, where you can select the
required serial protocol and proceed to the protocol setup. The key lights up if the
decoding of a serial bus is active.
If serial buses are active, pressing the key toggles these buses when the "Apps" dialog
or the "Serial Bus" dialog is open.
[FFT]
Opens and closes the "Math" dialog with FFT settings. The key lights up if an FFT is
active.
[Math]
Opens the "Math" dialog, where you can configure the calculation of mathematical
waveforms using various mathematic operations on other waveforms. Press the key
repeatedly to toggle the selected math waveform. If no math waveform is active, the
key closes the dialog.
If a math waveform is selected, the vertical rotary knobs are illuminated in light brown
(default color), the brightness of the color depends on the selected waveform.
[Position]
The vertical [Position] knob changes the vertical offset or the position of the selected
waveform. The horizontal axis and the selected waveform are moved vertically. You
can select whether the knob changes the offset or the position in the "Menu" > "Set-
tings" > "Frontpanel" > "Knobs" dialog.
●
Position indicates the vertical location in divisions.
●
Offset moves the vertical center of the selected channel to the offset value.
If the selected waveform is a math or reference waveform, serial bus, or logic channel,
the knob changes its vertical position.
The knob lights up in the color of the selected waveform. Turn clockwise to move up
the waveform. To set the value to zero, press the knob.
The current value is shown in the input box in the upper right corner of the screen.
Remote command:
Instrument tour