Remote control commands
R&S
®
RTO6
1729
User Manual 1801.6687.02 ─ 05
Alternatively, you can set the threshold according to the signal technology with
.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<TxThreshold>
User-defined clock threshold
Range:
-15 to 15
Increment: 0.1
*RST:
0
Default unit: V
BUS<m>:UART:TECHnology
<Technology>
Sets the threshold voltage Tx and Rx lines as defined for various signal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology>
V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V, 2.5 V, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MANual
Manual setting of user-defined values with
.
*RST:
V165
BUS<m>:UART:BITRate
<Bitrate>
Sets the number of transmitted bits per second.
Suffix:
<m>
.
1..4
Parameters:
<Bitrate>
Range:
300 to 20000000
Increment: 1
*RST:
9600
Default unit: bps
Usage:
Asynchronous command
Protocols