![R&S RTO6 Скачать руководство пользователя страница 596](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646596.webp)
Protocol analysis
R&S
®
RTO6
596
User Manual 1801.6687.02 ─ 05
"Start of frame" Searches for the first edge of the dominant SOF bit (synchronization
bit).
"Frame type"
Searches for a specified frame type (data, remote, error, or overload).
For data and remote frames, also the identifier format is considered.
For details, see:
●
●
"Identifier"
Searches for a specific message identifier or an identifier range.
See
"Identifier setup: Condition, Identifier min, Identifier max"
"Iden Data"
Searches for a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
The identifier conditions are the same as for the "Identifier" search cri-
teria, see
"Identifier setup: Condition, Identifier min, Identifier max"
on page 597. Data conditions are set with
Condition, Data min, Data max"
"Error condition"
Identifies various errors in the frame, see
"Symbolic"
The "Symbolic" search criteria is available if a DBC label list file is
loaded and applied. It allows you to search for specific data mes-
sages, or a signal and its value that appears inside the message.
Symbolic search is an alternative to the other search criteria, you can
either search for symbolic values or for an AND-combination of the
other 5 criteria.
For details, see
Chapter 13.7.6.3, "Symbolic search"
CAN and CAN FD (option R&S
RTO6-K520)