Triggers
R&S
®
RTO6
192
User Manual 1801.6687.02 ─ 05
● In the "Mode / Holdoff" tab, set the "Trigger mode" to "Normal".
6.2.2
Positioning the trigger
By positioning the trigger on the time axis, you define which part of the waveform is
displayed: mainly the pre-trigger part, or the post-trigger part, or the part around the
trigger point.
For details on position settings, see
Chapter 5.4, "Acquire settings"
1. Open the "Menu"> "Horizontal" dialog.
2. Set the "Reference point" and the "Position".
If you want to set the trigger position outside the waveform display, make sure that
"Restrict horizontal position to acquisition range" is disabled.
6.2.3
Using holdoff
For details on holdoff settings, see
1. Open the "Menu">"Trigger" dialog.
2. Select the "Mode / Holdoff" tab.
3. Select the "Holdoff mode".
4. Enter the "Holdoff settings" belonging to the selected mode.
6.2.4
Setting up an A
→
B
→
R trigger sequence
The complete configuration of a complex "A
→
B
→
R" trigger sequence consists of:
●
A-trigger condition
●
B-trigger condition in the same way as for the A-trigger, and optional delay time
between the two triggers
●
Optional reset by timeout and/or R-trigger
For details on sequence settings, see
1. Open the "Menu" > "Trigger" dialog.
2. Set "Trigger on" = "Sequence".
3. Select the type of "Sequence": "A
→
B
→
R".
4. Tap "Setup A trigger".
A dialog opens where you can configure the first event.
See:
Chapter 6.2.1, "Configuring a simple trigger"
5. Tap "Setup B trigger" and configure the B-trigger condition.
Setting up the trigger