![R&S RTO6 Скачать руководство пользователя страница 696](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646696.webp)
Protocol analysis
R&S
®
RTO6
696
User Manual 1801.6687.02 ─ 05
"Word Count Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:MILStd:CMD:CCONdition
T/R (Transmit/receive)
Toggles the data direction of the selected command: 1 (transmit), 0 (receive), or X
(either).
Remote command:
Status Flags
Specifies the values of the status flags. You can use the following characters: 1; 0; or X
(do not care).
For details, see
Remote command:
TRIGger<m>:MILStd:STATus:BCReceived
TRIGger<m>:MILStd:STATus:DBCaccept
TRIGger<m>:MILStd:STATus:INSTrument
TRIGger<m>:MILStd:STATus:MERRor
TRIGger<m>:MILStd:STATus:SREQuest
TRIGger<m>:MILStd:STATus:SUBSystem
TRIGger<m>:MILStd:STATus:TERMinal
Error Condition
Specify the error conditions to be triggered on.
Synchronization Error ← Error Condition
Triggers if a sync impulse does not fulfill the technical requirements or when the trans-
mission is not valid.
Remote command:
Manchester Coding Error ← Error Condition
Triggers if there is an error in the Manchester coding of the signal.
Remote command:
TRIGger<m>:MILStd:ERRor:MANChester
Parity Error ← Error Condition
Checks the parity of every word and triggers if the parity is even.
Remote command:
TRIGger<m>:MILStd:ERRor:PARity
MIL-1553 (option R&S
RTO6-K530)