![R&S RTO6 Скачать руководство пользователя страница 1020](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461020.webp)
Waveform generator (Option R&S
RTO6-B6)
R&S
®
RTO6
1020
User Manual 1801.6687.02 ─ 05
"Function type"
Min frequency
Max frequency
"Pulse"
0.001 Hz
30 MHz
"Cardinal sine"
0.001 Hz
5 MHz
"Cardiac"
0.001 Hz
1 MHz
"Gauss"
0.001 Hz
25 MHz
"Lorentz"
0.001 Hz
10 MHz
"Exp.rise"
0.001 Hz
10 MHz
"Ep. fall"
0.001 Hz
10 MHz
The values of the "Frequency" and "Period" depend on each other, as:
Period = 1/ Frequency
Remote command:
Period
Sets the period of the waveform. The available period range depends on the selected
"Function Type".
Remote command:
Duty cycle
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age of the period, the signal state is high.
Remote command:
WGENerator<m>:FUNCtion:SQUare:DCYCle
Symmetry
Sets the symmetry of a ramp waveform, the percentage of time the waveform is rising.
By changing the symmetry of the ramp, you can create, for example, triangular wave-
forms.
Remote command:
WGENerator<m>:FUNCtion:RAMP[:SYMMetry]
Pulse width
Sets the pulse width, the pulse duration of the generated pulse waveform.
Remote command:
WGENerator<m>:FUNCtion:PULSe[:WIDTh]
15.1.3
Modulation
Modulation is when properties of an original periodic waveform, the carrier signal, are
varied according to a second modulating signal. The type of modulation used deter-
mines which properties are changed.
Setup of the waveform generator