![R&S RTO6 Скачать руководство пользователя страница 884](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646884.webp)
Protocol analysis
R&S
®
RTO6
884
User Manual 1801.6687.02 ─ 05
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
For "Type">"HS Data":
TRIGger<m>:DPHY:DATA:CONDition
For "Type">"LP Escape Mode":
TRIGger<m>:DPHY:ESDTa:CONDition
Data Index
Data index sets the range within this series of the data value that is considered for the
analysis. The data index setup consists of the condition and one or two data index val-
ues.
"Condition"
Sets the comparison condition to a specific value or a range.
"Index Min/Index"
Specifies the value or sets the start value of a range.
"Index Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
For "Type">"HS Data":
TRIGger<m>:DPHY:DIDX:CONDition
For "Type">"LP Escape Mode":
TRIGger<m>:DPHY:ESINdex:CONDition
Escape mode
The escape mode setup consists of the condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
D-PHY (option R&S
RTO6-K580)