![R&S RTO6 Скачать руководство пользователя страница 1771](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461771.webp)
Remote control commands
R&S
®
RTO6
1771
User Manual 1801.6687.02 ─ 05
Setting parameters:
<Frame>
<Field>
<Enabler>
ON | OFF
Example:
:TRIGger:CMSB:ADVanced:FIEN "Master", "CRC", 1
Enables the CRC field of the Master frame.
:TRIGger:CMSB:ADVanced:FRAMe1:FLD3:ENABle 1
Enables the 3rd field of the 1st frame that you have defined.
Usage:
Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:BIT
<BitState>
TRIGger<m>:CMSB:ADVanced:BIT
<Frame>, <Field>, <Bit>
Sets the bit state of a field to be triggered on that only consists of one bit.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Bit>
ONE | ZERO | DC
ONE
1
ZERO
0
DC
"Do not care" (DC) = X
Usage:
Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMAX
<DataMax>
TRIGger<m>:CMSB:ADVanced:DMAX
<Frame>, <Field>, <Data>
Sets the end value of a data pattern range if
is set to
INRange
or
OORange
.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Field>
<Data>
Protocols