Triggers
R&S
®
RTO6
206
User Manual 1801.6687.02 ─ 05
Slope
Sets the edge type for the trigger condition.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
"Both"
Selects the rising as well as the falling edge. This option is not availa-
ble if the trigger source is the external trigger input.
Remote command:
Upper level
Sets the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:UPPer
Lower level
Sets the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:LOWer
Range
Selects how the time limit for the slew rate is defined. The time measurement starts
when the signal crosses the first trigger level - the upper or lower level depending on
the selected slope. The measurement stops when the signal crosses the second level.
Trigger types