Protocol analysis
R&S
®
RTO6
978
User Manual 1801.6687.02 ─ 05
Remote command:
EYE<m>:TIMReference:CLOCk:OFEStimation
Sampling time ← Explicit clock
Sets a sampling time for the explicit clock signal, an offset for the clock edge in relation
to the bit start. The clock edge sets the beginning of the unit interval. The sampling
time is a number between 0 and 1. Value 0 sets the clock edge to the beginning of the
bit period; value 0.5 sets the clock edge to the middle of the bit period.
Remote command:
EYE<m>:TIMReference:CLOCk:SAMTime
13.27.4.2
Display settings
Access: [App Cockpit]> "Analysis" > "DDR Eye" > "Display" tab
In the "Display" tab, you define the appearance of the individual eye diagrams.
You can configure and analyze up to four independent eye diagrams in parallel (vertical
subtabs Eye
1 to Eye
4). Make sure to select the correct eye diagram subtab on the
left before you adjust the settings.
Color table
Like any other waveform, you can show the eye waveform in solid color, or apply a
color table.
DDR (option R&S