![R&S RTO6 Скачать руководство пользователя страница 1896](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461896.webp)
Remote control commands
R&S
®
RTO6
1896
User Manual 1801.6687.02 ─ 05
Query parameters:
<SearchName>
Return values:
<Data>
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:FCSTate?
<SearchName>
Returns the status of the frame checksum.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<ChecksumState>
OK | STERror | SPERror | PRERror | UVAL | INSufficient
STERror: StarT ERror, incorrect start bit
SPERror: StoP ERror, incorrect stop bit
PRERror:PaRity ERror, incorrect parity bit.
UVAL: unexpected value
*RST:
OK
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:FCValue?
<SearchName>
Returns the value of the frame checksum.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<ChecksumValue>
Range:
0 to 16777215
Increment: 1
*RST:
0
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:FLAGs?
<SearchName>
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Protocols