Remote control commands
R&S
®
RTO6
1382
User Manual 1801.6687.02 ─ 05
Parameters:
<ScaleCoupling>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
DIFFerential<m>:COMMon:OFFSet
<Offset>
DIFFerential<m>:DIFFerential:OFFSet
<Offset>
Set the offset of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Offset>
Offset value
Range:
-1 to 1
Increment: 0.01
*RST:
0
Default unit: V
Usage:
Asynchronous command
DIFFerential<m>:COMMon:POSition
<Position>
DIFFerential<m>:DIFFerential:POSition
<Position>
Set the vertical position of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Position>
Position value
Range:
-5 to 5
Increment: 0.02
*RST:
0
Default unit: div
Usage:
Asynchronous command
24.8.10
Digital filter
...................................................................................... 1383
......................................................................................1383
Acquisition and setup